Loren Data's SAM Daily™

fbodaily.com
Home Today's SAM Search Archives Numbered Notes CBD Archives Subscribe
FBO DAILY ISSUE OF JANUARY 15, 2006 FBO #1511
MODIFICATION

J -- Fume Hood Face Velocity Readings, Inspections and Repairs

Notice Date
1/13/2006
 
Notice Type
Modification
 
NAICS
811310 — Commercial and Industrial Machinery and Equipment (except Automotive and Electronic) Repair and Maintenance
 
Contracting Office
Department of Commerce, National Institute of Standards and Technology (NIST), Acquisition Management Division, 100 Bureau Drive, Building 301, Room B129, Mail Stop 1640, Gaithersburg, MD, 20899-1640
 
ZIP Code
20899-1640
 
Solicitation Number
SB1341-06-Q-0116
 
Response Due
1/20/2006
 
Archive Date
2/4/2006
 
Point of Contact
Joni Laster, Purchasing Agent, Phone 301-975-8397, Fax 301-975-8884, - Joan Smith, Contract Specialist, Phone (301)975-6458, Fax (301)975-8884,
 
E-Mail Address
joni.laster@nist.gov, Joan.Smith@nist.gov
 
Description
Solicitation SB1341-06-Q-0116 is hereby amended to address vendor questions/answers. Question: In the Statement of Work, Background, it says "All interior positive pressure exhaust ducts shall be leak tested." In the Scope of Work it says "inspect the condition of ductwork for leaks". In neither case there is not mention of a procedure or reference a standard to be applied for this work. Is the intent for a visual/audible inspection of the ductwork for leakage or is there a procedure and criteria for performing leakage testing of the ductwork that will be provided? Answer: At this time only a visual / audible inspection of the entire system including the positive duct will be necessary. A more in depth inspection of the systems will be done in another phase of this work at a later date. The Contractor however, (at his discretion) may introduce a small amount of smoke or a specific odor (such as peppermint) to the system to aid in this inspection. The current response date remains unchanged. All responses shall be submitted not later than 3:30pm, local time on January 20, 2006.
 
Place of Performance
Address: 100 Bureau Drive,, Gaithersburg, Maryland
Zip Code: 20899
Country: USA
 
Record
SN00966296-W 20060115/060113211841 (fbodaily.com)
 
Source
FedBizOpps Link to This Notice
(may not be valid after Archive Date)

FSG Index  |  This Issue's Index  |  Today's FBO Daily Index Page |